Accuracy of working length determination with root ZX apex locator and radiography: An in vivo and ex vivo study

Vahid Zand1, Hadi Mokhtari1*, Mehrdad Lotfi1, Mohammad Forough Reyhani1, Aydin Sohrabi2, Pardis Tehranchi3 and Hamid Reza Mokhtari Zenouzi4
1Department of Endodontics, Faculty of Dentistry, Tabriz University of Medical Sciences, Tabriz, Iran. 2Department of Orthodontics, Faculty of Dentistry, Tabriz University of Medical Sciences, Tabriz, Iran. 3Department of Operative Dentistry, Faculty of Dentistry, Tabriz University of Medical Sciences, Tabriz, Iran. 4Department of Endodontics, Faculty of Dentistry, Zahedan University of Medical Sciences, Zahedan, Iran.
Email: [email protected]

  •  Accepted: 13 June 2011
  •  Published: 18 July 2011

Abstract

The purpose of this study was to clinically compare working length (WL)determination with root ZX apex locator and radiography, and then compare them with direct visualization method ex vivo. A total of 75 maxillary central and lateral incisors were selected. Working length determination was carried out using radiographic and electronic apex locator methods. Subsequently, the tooth under study was extracted and actual working length was determined directly under a stereomicroscope. Data were analyzed by Wilcoxon signed-rank, Spearman’s correlation coefficient and intra-class correlation tests. All the statistical analyses were set with a significance level of α = 0.05. The absolute measurement errors of the two methods were compared using Wilcoxon signed test, exhibiting no statistically significant difference in measurement errors between the two methods. Descriptive evaluation revealed that in 72% (n = 54) of the specimens, both methods had errors in the same direction and in 28% (n = 21) of the specimens, the two methods had errors in opposite directions. Intra-class correlation coefficient test demonstrated a high degree of agreement between the two methods. In conclusion, this study did not show any difference between radiography, root ZX and direct visualization in WL determination.

 

Key words: Working length, electronic apex locator, root ZX, radiography.