Full Length Research Paper

Performance of dried blood spot (DBS) PUNCHER and dried blood spots to measure HIV-1 viral load

Anoumou Dagnra
  • Anoumou Dagnra
  • Laboratoire BIOLIM, Faculté des Sciences de la Santé, Université de Lomé, Lomé, Togo.
  • Google Scholar
Mounerou Salou
  • Mounerou Salou
  • Laboratoire BIOLIM, Faculté des Sciences de la Santé, Université de Lomé, Lomé, Togo.
  • Google Scholar
Amivi Ehlan
  • Amivi Ehlan
  • Laboratoire BIOLIM, Faculté des Sciences de la Santé, Université de Lomé, Lomé, Togo.
  • Google Scholar
Ahouefa Konou
  • Ahouefa Konou
  • Laboratoire BIOLIM, Faculté des Sciences de la Santé, Université de Lomé, Lomé, Togo.
  • Google Scholar
Alassane Ouro-Medeli
  • Alassane Ouro-Medeli
  • Laboratoire BIOLIM, Faculté des Sciences de la Santé, Université de Lomé, Lomé, Togo.
  • Google Scholar
Sika Dossim
  • Sika Dossim
  • Laboratoire BIOLIM, Faculté des Sciences de la Santé, Université de Lomé, Lomé, Togo.
  • Google Scholar
Komlan Ali-Edje
  • Komlan Ali-Edje
  • Laboratoire BIOLIM, Faculté des Sciences de la Santé, Université de Lomé, Lomé, Togo.
  • Google Scholar


  •  Received: 17 July 2016
  •  Accepted: 06 September 2016
  •  Published: 21 September 2016

 ABSTRACT

The use of plasma is an obstacle to realize HIV-1 viral load in sub-Saharan Africa. In this context, the dried blood spot (DBS) is an interesting tool for sample collections. This approach was tested using a DBS hole-punch device (PUNCHER). Plasma and DBS samples were obtained from 102 patients, comprising 17 HIV-1 negative patients and 85 HIV-1 infected patients. The PUNCHER’s performance used to cut DBS was evaluated with the following criteria: ease of use, time savings and safety. VL was measured in parallel on plasma and DBS samples using NucliSENS EasyQ HIV-1. The correlation between plasma and DBS results was strong (R = 0.91; P < 0.001). The mean difference (± standard deviation) was -0.59 ± 0.52 log10 copies/ml. The sensitivity and specificity of DBS were 91.3% (n= 74) for the 81 VL detectable samples and 100% for the 21 VL undetectable samples, respectively. On a scale of 10, the PUNCHER’s performance scored 9.3 for ease of use, 8.6 for time savings and 10 for safety. PUNCHER is highly efficient at cutting DBS, and the VL resulting from DBS correlated well with those obtained from plasma.

Key words: Puncher, dried blood spot (DBS), viral load, performance.


 INTRODUCTION

In developed countries, viral load (VL) is an essential assay for monitoring the human immunodeficiency virus type 1 (HIV-1) infection, especially for evaluating the efficacy of antiretroviral treatment (ART) (Mellors et al., 1997; Yilmaz, 2001). In sub-Saharan Africa, the use of VL monitoring to detect treatment failure is the major challenge to improve HIV management. Access to this assay faces several obstacles: cost of equipment and reagents, availability and stability of energy. If VL measuring equipment exists in these countries, it is more often only available in reference laboratories far from peripheral sites, posing the problem of accessibility to people living with HIV. Then, the use of dried blood spot (DBS) represents an alternative for the collection and transport of samples as compared to plasma, which requires more restrictive conditions, transport without delay and storage at ‑80°C (Cassol et al., 1997; Brambilla et al., 2003; Alvarez-Muñoz et al., 2005; Kane et al, 2008 ; Johannessen,  et al 2009; Arredondo et al, 2012). In most laboratories, DBS are cut with a pair of scissors, with a risk of injury while cutting or decontaminating the scissors. In addition, discomfort from the long use of scissors limits the number of DBS that are cut.

The objectives of this study were, to evaluate the performance of the puncher, a new method of cutting DBS (weight = 4.6 kg; length = 48.5 cm; width = 21 cm; Figure 1; http://www.biomerieux-diagnostics.com) according to 7 criteria and to investigate the performance of DBS in HIV-1 RNA quantification against the standard plasma viral load assay.

 


 METHODS

A total of 102 patients, comprising 17 HIV-1 negative patients and 85 infected patients (58% of whom were antiretroviral therapy naive), were recruited at Sylvanus Olympio University Hospital in Lomé (Togo) over a period of six months, January to June 2013. After obtaining written informed consent, five milliliters of whole blood was drawn from each patient by venipuncture and collected in tubes with EDTA. DBS were prepared by dispensing 50 µl of blood per spot (5 spots per card) onto filter paper cards (Whatman no. 903; Schleicher and Schuell, BioScience GmbbH, Barcelona, Spain). The spotted filter papers were allowed to dry at room temperature for 4 to 6 h in a hood. The DBS were stored in zip-lock plastic bags with a silica gel desiccant at room temperature for 15 days before further processing and assaying. The remaining blood sample was centrifuged at 1500 xg and plasma was stored at -80°C until testing.

The DBS were cut with the PUNCHER according to the manufacturer’s instructions. During the use of the PUNCHER, the following performance-related parameters were evaluated: speed of handling, number of manipulations to cut the spots, speed of positioning the card; risk of contamination associated with handling. To avoid contamination of one sample by another, 3 white spots of the same blotting paper were cut after punching 2 spots of a DBS sample before moving on to the next sample. Seventeen (17) HIV-1 negative samples were included in the study to evaluate the contamination risk by using the PUNCHER. Criteria for PUNCHER evaluations were defined taking into account the difficulties associated with using scissors to cut DBS. The use of the PUNCHER was carried out by one lab technician. For each parameter, a score between 1 and 10 was awarded after each sample: 10 meaning completely in line with expectations. 

HIV-1 RNA isolation from DBS and plasma was performed using 100 µl (2 spots) and 500 µl of samples, respectively. HIV-1 RNA was extracted from the same patient in plasma and DBS according to the NucliSENS miniMAG procedure. VL was measured by NucliSENS HIV-1 EasyQ version 2.0 (bioMérieux, Lyon, France). All RNA values are reported as log10-transformed copy numbers of HIV RNA per ml of DBS or plasma. Viral load was stratified into three levels (undetectable VL, VL < 5000 copies/ml and VL > 5000 copies/ml). Sensitivity and specificity of DBS viral load using plasma assay as the gold standard was assessed at all three viral load strata. Pearson correlation analysis was performed, as well as Bland-Altman analysis to examine the level of agreement between the two tests (Bland et al., 1986). Bland-Altman analysis was performed using MedCalc version 9.5.0.0 (MedCalc Software, Mariakerke, Belgium). With the given sample size that was used for Bland Altman analysis, the 95% CI for the limits of agreement were ± 0.11 log10 copies/ml. This narrow range in the precision of the limits of agreement was deemed to be clinically acceptable. Differences were considered significant only when P values were <0.05. The Ministry of Health of Togo (No. 0411/2012/MS/CAB/ DGS/DPLET/CBRS) and National ethic committee of Togo (Comité de bioéthique pour la recherche en santé, No. 001/2012/CBRS) approved the study.


 RESULTS

The plasma results showed that all HIV-1 negative samples (n=17) were undetectable for VL results. From 85 infected patients, 81 (95.3%) were detectable VL (Table 1), including 24 VL < 5000 copies/ml and 57 VL > 5000 copies/ml. The correlation between DBS and plasma was high (R = 0.91; P < 0.001) (Figure 2). The mean difference and standard deviation in samples with a VL < 4 log10 (n = 30) was 0.65 (0.40); 0.60 (0.25) for samples between 4-5 log10 (n = 26) and 0.59 (0.15) for samples with a VL > 5 log10 (n = Bland-Altman plot illustrates the agreement between plasma and DBS (Figure 3). The mean difference for quantitative 25). Overall, 49 pairs of plasma and DBS (49%) had a difference of more than 0.5 log copy, and for 14 pairs of samples (14%), the difference was greater than 1 log10 copy. The corresponding data between measurements (DBS minus plasma) was -0.59 log10 copy (standard deviation, 0.52 log10 copies/ml). The sensitivity of DBS was 91.3% (n= 74) for the 81 VL detectable samples. Ten results from DBS were < 5000 copies/ml among patients who had VL > 5000 copies with plasma; for 6 of these 10 samples, the VL from plasma was between 5000 and 7000 copies/ml and for 4 samples the VL was > 11000 copies/ml. The specificity of DBS versus plasma was 100% for the 21 VL undetectable samples (17 HIV-1 negative samples and 4 HIV-1 positive samples).

 

 

 

 

On a scale of 10, PUNCHER scored 9.3 for ease of use, 8.6 for time savings (Table 2). Concerning the rapidity of cutting, 124 DBS corresponding to 62 patients were cut in one hour using the puncher. As compared to archives data (unpublished data), 88 DBS corresponding to 44 patients were cut using scissors in the laboratory. No contamination was observed, all 17 negative samples were found negative although randomly tested between samples with high viral loads. The absence of contamination proved the efficacy of the decontaminating protocol.

 


  DISCUSSION

The use of DBS specimen as source for diagnostic test has become increasingly popular in recent years. DBS has been used to identify genetic and metabolic disorders in neonates, detection of HIV-1 antibody, and HIV-1 DNA for infant diagnosis of HIV infection. The WHO recommends the use of DBS for HIV drug resistance surveillance for monitoring transmitted drug resistance in resources limited settings (WHO, 2010).

In this study, a new tool to cut DBS replacement scissors wa evaluated in order to validate its use in routine practice. As compared to scissors, the performance of the puncher concerning ease of use, time savings and rapidity of cutting was better. But the data obtained for the scissors are archive data. 

The VL results were divided into two groups based on the new WHO recommendations; patients on high active antiretroviral therapy for at least 4 weeks and with a VL > 5000 copies/ml are considered to be in treatment failure (WHO, 2012). The results showed a strong correlation between plasma and DBS. But the small sample size included in this study may limit the accuracy of the results; however, significant correlation and limits of agreement of two assay methods found in this study reinforce the usefulness and feasibility of utilizing DBS as method for clinical viral load monitoring of patients on ART or HIV-1 early diagnosis.

Previous data report correlations range from 78 to 99% between DBS and plasma according to different analyzers for the measurement of VL, and sometimes with different extraction methods (Brambilla et al., 2003; Johannessen et al., 2009; Garrido et al., 2009; Marconi et al., 2009; Mbida et al., 2009; Hamers et al., 2009; Bertagnolio et al., 2010; Johannessen et al., 2011b; Neogi et al., 2012). In this study, DBS were stored at room temperature (28 to 35°C) for 15 days before being handled, and were never frozen. 15 days was set as storage time at room temperature for DBS in this method to be used by the HIV care centers; which gives a time to transport DBS to the national reference laboratory.

HIV-1 RNA quantification from DBS has shown good stability under different temperature and storage conditions ranging from ambient to -70°C (Cassol et al., 1997; Brambilla et al., 2003; Alvarez-Muñoz et al., 2005; Kane et al., 2008; Marconi et al., 2009; Monleau et al., 2010). Nevertheless, subsequent studies should define the DBS VL values at which the patient will be considered either in remission or in virological failure.

In conclusion, the performance of the PUNCHER is excellent. It can therefore be recommended tolaboratories handling DBS to avoid the disadvantages and risks related to the use of a pair of scissors. The results is recommended also, because of its good performance for measuring viral load in tropical climatic conditions. DBS can be used as an alternative sampling method for viral load monitoring in resource-limited settings. Further studies of operational research to apply these findings within a clinical setting on a large scale will be useful.


 CONFLICT OF INTEREST

The authors have not declared any conflict of interest



 REFERENCES

Alvarez-Mu-oz MT, Zaragoza-Rodríguez S, Rojas-Montes O, Palacios-Saucedo G, Vázquez-Rosales G, Gómez-Delgado A, Torres J, Mu-oz O (2005). High correlation of human immunodeficiency virus type-1 viral load measured in dried-blood spot samples and in plasma under different storage conditions. Arch. Med. Res. 36(4):382-386.
Crossref

 

Arredondo M, Garrido C, Parkin N, Zahonero N, Bertagnolio S, Soriano V, de Mendoza C (2012). Comparison of HIV-1 RNA measurements obtained by using plasma and dried blood spots in the automated abbott real-time viral load assay. J. Clin. Microbiol. 50(3):569-572.
Crossref

 
 

Bertagnolio S, Parkin NT, Jordan M, Brooks J, García-Lerma JG (2010). Dried blood spots for HIV-1 drug resistance and viral load assaying: A review of current knowledge and WHO efforts for global HIV drug resistance surveillance. AIDS Rev. 12(4):195-208

 
 

Brambilla D, Jennings C, Aldrovandi G, Bremer J, Comeau AM, Cassol SA, Dickover R, Jackson JB, Pitt J, Sullivan JL, Butcher A, Grosso L, Reichelderfer P, Fiscus SA (2003). Multicenter evaluation of use of dried blood and plasma spot specimens in quantitative assays for human immunodeficiency virus RNA: measurement, precision, and RNA stability. J. Clin. Microbiol . 41(5):1888-1893.
Crossref

 
 

Bland JM, DG (1986). Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1(8476):307-310.
Crossref

 
 

Cassol S, Gill MJ, Pilon R, Cormier M, Voigt RF, Willoughby B, Forbes J (1997). Quantification of human immunodeficiency virus type 1 RNA from dried plasma spots collected on filter paper. J. Clin. Microbiol. 35(11):2795-27801.

 
 

Garrido C, Zahonero N, Corral A, Arredondo M, Soriano V, de Mendoza C (2009). Correlation between human immunodeficiency virus type 1 (HIV-1) RNA measurements obtained with dried blood spots and those obtained with plasma by use of Nuclisens EasyQ HIV-1 and Abbott RealTime HIV load assays. J. Clin. Microbiol. 47(4):1031-1036
Crossref

 
 

Hamers RL, Smit PW, Stevens W, Schuurman R, Rinke de Wit TF (2009). Dried fluid spots for HIV type-1 viral load and resistance genotyping: a systematic review. Antivir. Ther. 14(5):619-629.

 
 

Johannessen A, Garrido C, Zahonero N, Sandvik L, Naman E, Kivuyo SL, Kasubi MJ, Gundersen SG, Bruun JN, de Mendoza C (2009). Dried blood spots perform well in viral load monitoring of patients who receive antiretroviral treatment in rural Tanzania. Clin. Infect. Dis. 49(6):976-981.
Crossref

 
 

Johannessen A, Garrido C, Zahonero N, Naman E, de Mendoza C (2011). HIV-1 drug resistance assaying from dried blood spots collected in rural Tanzania using the ViroSeq HIV-1 Genotyping System. J. Antimicrob. Chemother. 66(2):260-264
Crossref

 
 

Kane CT, Ndiaye HD, Diallo S, Ndiaye I, Wade AS, Diaw PA, Gaye-Diallo A, Mboup S (2008). Quantitation of HIV-1 RNA in dried blood spots by the real-time NucliSENS EasyQ HIV-1 assay in Senegal. J. Virol. Methods 148(1-2):291-295.
Crossref

 
 

Marconi A, Balestrieri M, Comastri G, Pulvirenti FR, Gennari W, Tagliazucchi S, Pecorari M, Borghi V, Marri D, Zazzi M (2009). Evaluation of the Abbott Real-Time HIV-1 quantitative assay with dried blood spot specimens. Clin. Microbiol. Infect. 15(1):93-97
Crossref

 
 

Mbida AD, Sosso S, Flori P, Saoudin H, Lawrence P, Monny-Lobé M, Oyono Y, Ndzi E, Cappelli G, Lucht F, Pozzetto B, Oukem-Boyer OO, Bourlet T (2009). Measure of viral load by using the Abbott Real-Time HIV-1 assay on dried blood and plasma spot specimens collected in 2 rural dispensaries in Cameroon. J. Acquir. Immune. Defic Syndr. 52(1):9-16
Crossref

 
 

Mellors JW, Mu-oz A, Giorgi JV, Margolick JB, Tassoni CJ, Gupta P, Kingsley LA, Todd JA, Saah AJ, Detels R, Phair JP, Rinaldo CR Jr (1997). Plasma viral load and CD4+ lymphocytes as prognostic markers of HIV-1 infection. Ann. Intern. Med. 126(12):946-954.
Crossref

 
 

Monleau M, Butel C, Delaporte E, Boillot F, Peeters M (2010). Effect of storage conditions of dried plasma and blood spots on HIV-1 RNA quantification and PCR amplification for drug resistance genotyping. J. Antimicrob. Chemother. 65(8):1562-1566.
Crossref

 
 

Neogi U, Gupta S, Rodridges R, Sahoo PN, Rao SD, Rewari BB, Shastri S, Costa AD, Shet A (2012). Dried blood spot HIV-1 RNA quantification: a useful tool for viral load monitoring among HIV-infected individuals in India. Indian J. Med. Res. 136(6):956-962.

 
 

WHO (2010). Technical Brief on HIV viral load technologies. Available at: View.

 
 

WHO (2010). Manel for HIV drug resistance testing using dried blood spot specimen, March 2010. Available at: 

View.

 
 

WHO (2012). Antiretroviral therapy for HIV infection in adults and adolescents : recommendations for public health approach.-2010 rev. Available at: 

View, accessed on June 13 2012

 
 

Yilmaz G (2001). Diagnosis of HIV infection and laboratory monitoring of its therapy. J. Clin. Virol. 21(3):187-196.
Crossref

 

 




          */?>